Gold Cluster Labeling Calculation: Advanced Method
|
|
Updated: September 30, 2007
This calculation method and examples are taken from workshop, "Gold Cluster Labeling and STEM Microscopy" held at Brookhaven National Laboratory, September 18-21, 1996. In this section:Calculation method
|
The extinction coefficients for the undecagold and Nanogold® clusters have been determined at 280 nm and 420 nm. These values are shown below, and also the values for IgG molecules and Fab' fragments.
Calculation of extinction coefficients from optical densities
Note: The actual value for the extinction coefficient of Nanogold at 280 nm varies slightly from lot to lot. The most accurate value for the lot you are using will be given in the product specification sheet supplied with your product: use this value for the most accurate labeling calculation.
Absorbances of proteins are often given as optical densities (ODs), which expresses absorbance as a funtion of the mass of protein dissolved per unit volume. Usually the literature values (E1%
Calculation of gold cluster labeling
Where both the protein and the gold cluster absorb at 280 nm and 420 nm, the observed absorbance at each wavelength is expressed as a pair of simultaneous equations:
Where:
c1 = measured absorbance at lambda = 280 nm
Applying Cramer's Rule and solving for x and y in terms of a, b, and c gives:
This gives the general solution shown below.
General solution
Labeling of IgG and Fab'
With antibody labeling, the calculation is simplified because these proteins do not absorb at 420 nm; therefore, a2 = 0. This gives the simplified solution below:
Simplified Solution
The following values were measured for UV/visible absorbtion:
280 nm: 0.6524
Sample volume: 3.2 mL.
For this sample, the values of the constant terms are:
c1 = 0.6524 (measured absorbance at lambda = 280 nm)
(ii) Hemoprotein labeled with undecagold:
A hypothetical heme-containing protein with a moleular weight of 65,000 and extinction coefficients of 37,400 at 280 nm and 27,900 at 420 nm is labeled with mono-Sulfo-NHS-undecagold. 2.0 mL of labeled conjugate solution is isolated and absorbances were measured:
280 nm: 0.8705
Values for the constants are:
c1 = 0.8705 (measured absorbance at lambda = 280 nm)
These results show 1 undecagold cluster for every 5 heme-proteins; this labeling might indicate that the protein forms pentamers, each of which accomodates one gold label - STEM can then be used to investigate further.
c2 = measured absorbance at lambda = 420 nm
a1 = extinction coefficient of protein at lambda = 280 nm
a2 = extinction coefficient of protein at lambda = 420 nm
x = concentration of protein (moles/liter)
b1 = extinction coefiicient of gold cluster label at lambda = 280 nm
b2 = extinction coefficient of gold cluster label at lambda = 420 nm
y = concentration of gold cluster label (moles/liter)
Examples
(i) Fab' labeled with Nanogold®
420 nm: 0.2012
c2 = 0.2012 (measured absorbance at lambda = 420 nm)
a1 = 7.5 X 104 (extinction coefficient of Fab' at lambda = 280 nm)
a2 = 0 (extinction coefficient of Fab' at lambda = 420 nm)
x = concentration of protein (moles/liter)
b1 = 4.12 X 105 (extinction coefiicient of Nanogold® at lambda = 280 nm)
b2 = 1.56 X 105 (extinction coefficient of Nanogold® at lambda = 420 nm)
420 nm: 0.4560
c2 = 0.4560 (measured absorbance at lambda = 420 nm)
a1 = 3.74 X 104 (extinction coefficient of protein at lambda = 280 nm)
a2 = 2.79 X 104 (extinction coefficient of protein at lambda = 420 nm)
x = concentration of protein (moles/liter)
b1 = 1.68 X 105 (extinction coefiicient of undecagold at lambda = 280 nm)
b2 = 4.71 X 104 (extinction coefficient of undecagold at lambda = 420 nm)